Appendix A: Exposure value tables for KaVo ProXam
iX
A.1 Default exposure values
Adult
Child
A.2 Preprogrammed settings values
Short cone 20 cm (8")
Maxilla
Adult
Mandible
Maxilla
Child
Mandible
Maxilla
Adult
Mandible
Maxilla
Child
Mandible
Long cone 30 cm (12")
Adult
Maxilla
Mandible
Child
Maxilla
Mandible
Maxilla
Adult
Mandible
Instructions for use
The following table shows the default exposure values for KaVo ProXam iS
with no target selected.
Short cone
kV
mA
63
8
60
8
INCISORS
kV
mA
60
8
60
8
OCCLUSAL
kV
mA
70
8
66
8
INCISORS
kV
mA
60
8
60
8
OCCLUSAL
kV
mA
70
8

Appendix A: Exposure value tables for KaVo ProXam iX

s
0,1
Adult
0,08
Child
PREMOLARS AND
CANINES
time
kV
mA
0.080
63
8
0.063
0.063
60
8
0.050
ENDODONTIC
time
kV
mA
0.08
60
8
0.063
60
8
PREMOLARS AND
CANINES
time
kV
mA
0.16
63
8
0.125
0.125
60
8
0.1
ENDODONTIC
time
kV
mA
0,16
60
8
Long cone
kV
mA
63
8
60
8
MOLARS
time
kV
mA
0.1
63
0.08
0.08
60
0.063
BITE-WING
time
kV
mA
0.08
60
0.063
60
MOLARS
time
kV
mA
0.2
63
0.16
0.16
60
0.125
BITE-WING
time
kV
mA
0.16
60
KaVo ProXam iS
s
0,2
0,16
time
0.125
8
0.1
0.1
8
0.08
time
8
0.125
8
0.1
time
0.25
8
0.2
0.2
8
0.16
time
8
0.25
23