Risk of injury and risk of product damage
► Comply with the product's field of application and do not expose
it to excessive strain (see page 6).
► Note the combination possibilities/combination exclusions in the
instructions for use of the products.
► Observe the maximum lifetime of the product.
NOTICE!
Risk of product damage and limited functionality
► Do not expose the product to prohibited environmental condi­
tions.
► Check the product for damage if it has been exposed to prohib­
ited environmental conditions.
► Do not use the product if it is damaged or in a questionable con­
dition. Take suitable measures (e.g. cleaning, repair, replace­
ment, inspection by the manufacturer or a specialist workshop).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an
altered gait pattern, a change in the positioning of the prosthetic com­
ponents relative to each other and by the development of noises.
4 Scope of delivery
Quantity
Designation
1

Instructions for use

1
Adapter
4
Set screw (only 6A54)
Spare parts/accessories (not included in the scope of deliv­
Designation
Set screw – 12 mm
Set screw – 14 mm
Set screw – 16 mm
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec­
tions
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and
about how to secure the screws.
Reference
5.1 Adjusting adapter
number
The adapter includes a pyramid adapter adjustable in 2 mm incre­
ments. The pyramid adapter can be aligned proximally or distally.
Required materials: 710D20 torque wrench, 636K13 Loctite 241
>
506G3=M8x12
1) Align the sliding adapter in the desired displacement direction
(a – p or m – l).
2) Lift the ring on the adjustable pyramid adapter slightly in order to
displace the pyramid adapter (see fig. 3).
ery)
Reference num­
ber
506G3=M8X12-V
506G3=M8X14
506G3=M8X16