Figure 3-1: Service UI menu

Individual Calibrations
Execute All Calibrations
AWG5000B and AWG5000C Series Service Manual
The left half of the window with the title Calibration is the calibration part of the
window. You can perform the calibrations either all in one step or individually.
However, since the calibration sequence is predefined, follow the sequence from
the top. You must finish a calibration item listed on the upper part of the screen
before performing the lower part of the items. Some items may be grayed out,
depending on your instrument option configurations.
To perform individual calibrations, click the calibration buttons, starting with
10 MHz Reference..., and the corresponding calibrations are executed.
To perform all calibrations in one step, click Execute All Calibrations and all
calibrations are executed in order starting from the top.
Adjustment Procedures
3–3