Levelflex FMP55 FOUNDATION Fieldbus
Endress+Hauser
12.8.1
Event groups
The diagnostic events are divided into 16 groups according to the source and the
weighting. A default event category is assigned to each group at the factory. Here, one
bit of the assignment parameters belongs to every event group.
Event weighting
Default event category
Highest weighting
Failure (F)
1)
This event can be removed from the group and treated separately; see the "Configurable area" section.
Event weighting
Default event category
High weighting
Function check (C)
Event weighting
Default event category
Low weighting
Out of specification (S)
Commissioning (block-oriented operation)
Event source
Bit
Events in this group
Sensor
31
• F003: Broken probe detected
• F046: Buildup detected
• F083: Memory content
• F104: HF cable
• F105: HF cable
• F106: Sensor
Electronics
30
• F242: Software incompatible
• F252: Modules incompatible
• F261: Electronic modules
• F262: Module connection
• F270: Main electronic failure
• F271: Main electronic failure
• F272: Main electronic failure
• F273: Main electronic failure
• F275: I/O module failure
• F276: I/O module failure
• F282: Data storage
• F283: Memory content
• F311: Memory content
Configuration
29
• F410: Data transfer
• F411: Upload/download
• F435: Linearization
• F437: Configuration incompatible
Process
28
• F803: Loop current 1
• F825: Operating temperature
• F936: EMC interference
• F941: Echo lost
• F970: Linearization
Event source
Bit
Events in this group
Sensor
27
Not used in Levelflex
Electronics
26
Not used in Levelflex
Configuration
25
• C411: Upload/download
• C431: Trim
• C484: Failure mode simulation
• C485: Simulation measured value
• C491: Simulation current output
• C585: Simulation distance
Process
24
Not used in Levelflex
Event source
Bit
Events in this group
Sensor
23
Not used in Levelflex
Electronics
22
Not used in Levelflex
1)
85
Need help?
Do you have a question about the Levelflex FMP55 and is the answer not in the manual?