Levelflex FMP55 FOUNDATION Fieldbus
Switch status
Navigation
Prerequisite
Description
Selection
Additional information
Simulation device alarm
Navigation
Description
Selection
Additional information
Diagnostic event simulation
Navigation
Description
Additional information
Endress+Hauser
Expert → Diagnostics → Simulation → Switch status
Switch output simulation (→ 200) = On
Select the status of the status output for the simulation.
• Open
• Closed
The switch status assumes the value defined in this parameter. This helps to check correct
operation of connected control units.
Expert → Diagnostics → Simulation → Sim. alarm
Switch the device alarm on and off.
• Off
• On
When selecting the On option, the device generates an alarm. This helps to check the
correct output behavior of the device in the case of an alarm.
An active simulation is indicated by the C484 Simulation failure mode diagnostic
message.
Expert → Diagnostics → Simulation → Diag. event sim.
Select a diagnostic event to simulate this event.
When operated via the local display, the selection list can be filtered according to the event
categories (Diagnostic event category parameter).
Operating menu
201
Need help?
Do you have a question about the Levelflex FMP55 and is the answer not in the manual?
Questions and answers