Operating menu
Description
Additional information
Enter access code
Navigation
Description
User entry
Additional information
140
Indicates access authorization to parameters via local display.
The access authorization can be changed via the Enter access code parameter
(→  140).
If additional write protection is active, this restricts the current access authorization
even further. The write protection status can be viewed via the Locking status
parameter (→  139).
Setup → Advanced setup → Ent. access code
Enter access code to disable write protection of parameters.
0 to 9 999
• The customer-specific access code that was defined in the Define access code parameter
(→  180) must be entered for local operation.
• If an incorrect access code is entered, users retain their current access authorization.
• The write protection affects all parameters marked with the
On the local display, the
write-protected.
• If no key is pressed for 10 minutes or the user goes from the navigation and editing
mode back to the measured value display mode, the device automatically locks the write-
protected parameters after another 60 s.
Please contact your Endress+Hauser Sales Center if you lose your access code.
Levelflex FMP55 FOUNDATION Fieldbus
symbol in front of a parameter indicates that the parameter is
symbol in the document.