Operating menu
Switch-on delay
Navigation
Prerequisite
Description
User entry
Switch-off value
Navigation
Prerequisite
Description
User entry
Additional information
168
A
Switch-on value
B
Switch-off value
C
Output closed (conductive)
D
Output opened (non-conductive)
 Setup → Advanced setup → Switch output → Switch-on delay
• Switch output function (→  165) = Limit
• Assign limit (→  166) ≠ Off
Define delay for the switch-on of status output.
0.0 to 100.0 s
 Setup → Advanced setup → Switch output → Switch-off value
Switch output function (→  165) = Limit
Enter measured value for the switch-off point.
Signed floating-point number
The switching behavior depends on the relative position of the Switch-on value and
Switch-off value parameters; description: see the Switch-on value parameter
(→  167).