Operating menu
Empty calibration
Navigation
Description
User entry
Factory setting
Additional information
Full calibration
Navigation
Description
128
The Medium group parameter presets the Medium property parameter as follows:
Medium group
Others
Water based (DC >= 4)
The Medium property parameter can be changed at a later point of time. However,
when doing so, the Medium group parameter retains its value. Only the Medium
property parameter is relevant for the signal evaluation.
The measuring range may be reduced for small dielectric constants. For details refer
to the Technical Information (TI) of the respective device.
Setup → Empty calibr.
Distance process connection to min. level.
Depending on the probe
Depending on the probe
33
Empty calibration (E) for interface measurements
In the case of interface measurements the Empty calibration parameter is valid for
both, the total and the interface level.
Setup → Full calibr.
Span: max. level - min level.
Levelflex FMP55 FOUNDATION Fieldbus
Medium property
Unknown
DC 4 ... 7
E
0%
E
0%
Endress+Hauser
A0013177