Filter Replacement Procedure - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
8-2-5

Filter Replacement Procedure

8-2-5-1
Manpower
One person, 30 minutes
8-2-5-2
Tools
No special tools are needed
Filter Cover
8-2-5-3
Filter Replacement Procedure
1.) Remove Filter Cover as described in
2.) Pull out plastic plugs and remove the filter.
Two different Plastic plugs have been used, the one to the left on early Vivid 7 scanners and the
type to the right on newer units (both Vivid 7 and Vivid 7 PRO).
3.) Install a new or a cleaned filter (GE Vingmed Part No: FB307903).
Use plastic plugs shown to the right (GE Vingmed Part No: 080X6520)
Replace this plug...
4.) Install Filter Cover as described in
02
EVISION
Filter
Figure 8-6 Filter Cover and Filter
Figure 8-7 Plug replacement
Chapter 8 - Replacement Procedures
"Filter Cover Removal Procedure" on page
with this plug
"Filter Cover Installation Procedure" on page
V
7 / V
7 PRO S
IVID
IVID
ERVICE
Plastic plugs
8-6.
8-6.
M
ANUAL
8 - 7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 7 proFc091194

Table of Contents