Tissue Velocity Imaging (TVI) Checks

4-3-8-1
Introduction
TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by
sampling of tissue Doppler velocity values at discrete points.
4-3-8-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner's left-most probe connector.
-
-
2.) Turn ON the scanner
-
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display's Brightness and Contrast setting.
4.) Press TVI.