AC Transformer Removal Procedure
Screw A
Ratcover
Old expansion plug
1.) Loosen screw A, B and C (See
Transformer to prevent it from tilting forwards.
2.) Push AC Transformer to the left, tilt it forwards to free screw keyholes from screw heads (A, B and
C) and lift the transformer off hooks along its the lower inner edge.
8-4-1-5
AC Transformer Installation Procedure
1.) Hang the inner lower edge of AC Transformer in its hooks.
2.) Adjust its position so that screw keyholes matches screw heads (A,B and C).
3.) Tilt AC Transformer upwards and push it to the right, screw heads into the keyholes.
4.) Hold it steady while you fasten screw D and E.
5.) Fasten screw A, B and C.
6.) Connect two cables.