8-4-2
AC Power Replacement Procedure
8-4-2-1
Manpower:
One person, 1 hour.
8-4-2-2
Tools:
Phillips screwdriver size 2
Dangerous voltages are present in this equipment. Completely power off and
DANGER
shut down the scanner before servicing.
8-4-2-3
Preparations
1.) Shut down the system as described in
2.) Remove Right Side Cover as described in
3.) Remove Filter Cover as described in
4.) Remove Lower Rear Cover as described in
5.) Note Voltage type of AC Power Module being replaced (see label on rear of system).
8-4-2-4
AC Power Removal Procedure
1.) Disconnect Speed Control cable, Fan cable and Input Voltage cable.
2.) Remove four (4) screws holding AC Power.
3.) Turn AC Power Module to get better access and unplug the rest of the cables. Make sure each
connector and socket are properly marked, and easy to identify during installation.
Fan Cable
8 - 34
02
EVISION
Peripheral Voltage Select
Speed Cable
Input Voltage cable
Figure 8-28
AC Power Module
Section 8-4 - FRU Replacement Procedures
V
"Power Shut Down" on page
"Side Covers Removal Procedure" on page
"Filter Cover Removal Procedure" on page
"Lower Rear Cover Removal Procedure" on page
Box turned for access to cabling.
Screws
7 / V
7 PRO S
IVID
IVID
ERVICE
4-5.
8-6.
M
ANUAL
8-3.
8-8.