GE Vivid 7 Service Manual page 130

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
3-9-3-7-8
Setting the DICOM Server as Default Dataflow
1.) Finnish the verification in
You need to have green check marks on both Hardware Ping and DICOM Ping before you continue.
1.) Click the DATAFLOW tab.
2.) Select "DICOM Server" dataflow on the "Dataflow Name" drop-down menu.
.
DATAFLOW tab
3.) Check the Default box.
.
4.) Exit Config and restart the scanner (by hitting CTRL-ALT-R at the same time).
3 - 54
02
EVISION
"Check the Connection to the DICOM Server (Ping)" on page
Figure 3-59 Selecting DICOM Dataflow Services
Figure 3-60 Set DICOM as Default Dataflow
Section 3-9 - Connectivity
V
7 / V
7 PRO S
IVID
IVID
ERVICE
Dataflow Name
drop-down menu
Select
"DICOM Server"
Click once in this check box
to make DICOM the Default
dataflow
M
ANUAL
3-53.