Acquisition Diagnostics-Test All
One test is available, Test All.
Test All runs a sequential test of the following;
1.) PC2IP
2.) IMP (Image Port Board)
3.) VME (VME data bus on the Back Plane)
4.) SDP (Spectrum Doppler Processor)
5.) RFT (Radio Frequency & Tissue Board)
6.) FEC (Front End Controller Board) (Some parts of TX board is also tested)
7.) BF (Beam Former Boards)
8.) RX (Receiver Board)
9.) TXP (TX Power)
NOTE:
Relay Board (RLB) and Chain are not tested in this test. Some parts of the Transmitter (TX)
Board is tested during the FEC Board test, but other parts of the TX Board are not tested at all.
7-2-7-2
Instructions Frame
Displays a test-specific page or the default instructions page.
7-2-7-3
Status Frame
Initially displays the last known status for the particular diagnostic. Once the diagnostic starts, the frame
displays the "current" status of all test results.