GE Vivid 7 Service Manual page 407

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
8-3-10-2
Application SW Installation
1.) Insert the applicable CD labelled Vivid7 & EchoPAC PC Application SW, according to
on page 8-25,
Table 8-13
SOFTWARE
SW. v.2.2.0
SW. v.1.1.3
2.) Hold down the power switch on the operator panel for a few seconds to power down the scanner
(forced shut down).
3.) Wait until the LCD's on the Operator Panel is switched OFF and the Power Switch on the Operator
Panel has turned yellow.
4.) Power ON the System as described in
5.) A dialogue box, called Start Application, will appear, select INSTALL SW.
6.) A dialogue box, called Startloader, will appear, select OK to continue.
7.) A dialogue box will appear, also named Startloader stating "New System SW installation will start
now", select OK to continue.
8.) A Command prompt window will appear, hit any key on the alphanumeric keyboard to continue
9.) A message in the Command prompt window may prompt you that copying application software
takes 2 minutes.
THIS WILL TAKE APPR. 5 MINUTES.
10.)Several dialogue boxes will appear on the screen during the SW load.
Do not interrupt the SW loading at any time.
NOTICE
11.)Remove the CD from the CD drive when prompted.
12.)The Scanner will now power down and reboot. Verify that the command prompt window is selected.
02
EVISION
into the CD drive.
Application Software List
DESCRIPTION
Vivid7 & EchoPAC PC
Application SW
(BT'02-M4-2)
Vivid7/EchoPAC PC Application
SW
(BT'01-M4)
Chapter 8 - Replacement Procedures
V
IVID
PART NUMBER
FC200220-04
FC200090-06
Table 8-5 on page
8-11.
7 / V
7 PRO S
IVID
ERVICE
Table 8-13
M
ANUAL
8 - 25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 7 proFc091194