Safety Considerations

1-3-1

Introduction

The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2

Human Safety

Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in a Vivid 7 / Vivid 7 PRO Training Seminar are authorized to
service the equipment.
1-3-3

Mechanical Safety

WHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE
WARNING
WARNING
EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
WARNING
WARNING
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
WARNING
WARNING
MAY BE DAMAGED.
Ensure that nobody touch the console arm/frogleg when moving the keyboard console.
CAUTION
Do not move the unit if the keyboard console is in unlocked position.
CAUTION
Always lock the Control Console in its parking (locked) position before moving the scanner
CAUTION
around.
WHEN THE TOP CONSOLE IS IN ITS LOCKED POSITION, THE GAS SPRING IS
WARNING
WARNING
COMPRESSED AND STORES MECHANICAL ENERGY. DURING NORMAL OPERATION
THE TOP CONSOLE, THE WEIGHT OF THE MONITOR AND THE MECHANICAL FORCE
OF THE GAS SPRING ARE IN BALANCE. TAKE CARE IF/WHEN YOU ACTIVATE THIS
GAS SPRING. PERSONAL INJURY CAN OCCUR AFTER THE PANEL IS REMOVED AND
THE SPRING PRESSURE IS RELEASED. TAKE CARE WHEN YOU REPAIR THE
ELEVATION ASSEMBLY.