GE Vivid 7 Service Manual page 367

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
7-3-3-3
Monitor Test
Double-click Monitor Test from the PC Diagnostic interactive tests menu.
The monitor test dialog box presents you with three options:
Combination Test - Checks the monitor for alignment, color depth, and monitor resolution. It will
put four cross-hairs in each corner, RGB (Red Green Blue) spectrum arrays and horizontal and
vertical alignment bars on the screen.
NOTE:
In most cases, you will choose the Combination Test.
02
EVISION
Figure 7-49 Monitor Test
Figure 7-50 Monitor Test
Chapter 7 - Diagnostics/Troubleshooting
V
7 / V
7 PRO S
IVID
IVID
ERVICE
M
ANUAL
7 - 55