GE Vivid 7 Service Manual page 397

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
8-3-7
Prepare MO Disk for Patient Archive
NOTE:
A separate MO disk is required for backup of Patient Database, System Configuration and User
Defined Configurations (customer presets).
This MO disk must be a new disk with no images stored.
1.) Label the MO Disk sequentially.i.e Backup 1A (side A), Backup 1B (side B).
2.) If the MO Disks is already formatted and labelled move to
NOTE:
When formatting the MO disk, give it a logical name and make sure to physically label the disk.
A labelling convention should be followed so that each MO disk gets a unique label. The
maximum numbers of label characters allowed is 11.
MO Disk Label side B must differ from side A.
3.) Insert a MO disk, side A, into the empty MO drive.
4.) Select CONFIG (F2) on the alphanumeric Keyboard on the operator panel.
5.) From the Operator Login window select ADM from the Operator pull-down menu.
6.) Select LOG ON.
7.) Select CONNECTIVITY in the Image and Analysis menu.
8.) Select TOOLS in the Connectivity menu.
9.) In the Label window, type the same name as labelled on the MO disk side A.
10.)Select FORMAT.
11.)A warning "Are you sure you want to format H: and remove all data from the media" appears, select
OK.
12.)Wait until the information dialogue window appears on screen.
13.)Select OK to continue.
02
EVISION
Figure 8-1 Prepare MO Disk - Patient Archive
Chapter 8 - Replacement Procedures
V
7 / V
7 PRO S
IVID
IVID
section 8-3-8 on page 8-16
M
ERVICE
ANUAL
8 - 15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 7 proFc091194

Table of Contents