S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
Section 7-4

Acquisition Diagnostics

7-4-1

System Test Diagnostics Summary

If the scanner is failing during start up or it don't start at all, please refer to
Boot
If the scanner is running, but an error that may be a hardware failure, has been registered, then
continue to read.
7-4-2

Start System Test

The Config Screen gives you access to the System Test (an automatic test).
To start the System Test, switch the scanner on, see
instructions.
After scanner initialization is complete, continue with the steps below;
1.) Press CONFIG on the keyboard and log on as adm to the System Config Screen.
2.) Select SYSTEM from the bottom of the menu that appears on the monitor.
3.) If not selected, click TEST.
.
4.) Click GO TO SYSTEMTEST to select the Calibration/Test screen.
02
EVISION
on page 7-67.
Figure 7-54 System > Test Selection Screen
Chapter 7 - Diagnostics/Troubleshooting
V
7 / V
IVID
IVID
4-2-1 - Power ON/ Boot Up
7 PRO S
M
ERVICE
ANUAL
7-5-1-1 - System Doesn't
on page
4-2
for specific
7 - 59