GE Vivid 7 Service Manual page 417

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
8-4-2-5
AC Power Installation Procedure
Before installing a replacement unit, verify that the power selector on the AC
DANGER
Power Module is set to the correct voltage for the peripherals used on the
system.
1.) Connect the cables. You may need to leave the Speed Control cable and the Fan cable
unconnected until you have fastened the unit. Input Voltage connector and socket are coded to
prevent mistakes. In 115V system the connector matches one end of the 8-holed socket. In 230V
system it matches the other end of the socket.
can only be plugged in the right end of the socket.
2.) Fasten AC Power Module for correct Voltage and connect the remaining cables.
3.) Install Lower Rear Cover as described in
4.) Install Filter Cover as described in
5.) Install Right Side Cover as described in
6.) Do functional check-out as described in Chapter 4..
Red blocks prevents wrong connections
02
EVISION
8 holed socket
Figure 8-29 Input Voltage connector/socket
Chapter 8 - Replacement Procedures
Figure 8-29
"Lower Rear Cover Installation Procedure" on page
"Filter Cover Installation Procedure" on page
"Side Covers Installation Procedure" on page
4 pin connector
V
7 / V
7 PRO S
IVID
IVID
ERVICE
is on a 230V system, and the connector
8-6.
M
ANUAL
8-8.
8-4.
8 - 35