Restore Standard Printer IP and Port Settings (cont'd)
9.) Verify Printer Name or IP Address as recorded in
10.)Select Raw.
11.)Verify Port Number as recorded in
12.)Select OK.
Printer Name or IP
Raw
Port Number
13.)Close rest of the Windows by selecting OK.
8-3-10-5
DC Offset Alignment
1.) From the upper section of the window, select TEST.
2.) Then select GO TO SYSTEM TEST.
3.) From the Module selection block, select DC OFFS. CAL.
4.) Reboot the System.
8-3-10-6
Verify System Software Version
1.) From the alphanumeric keyboard on the Operator Panel, select CONFIG (F2 key), from the
operator log in window, select ADM from the pull down menu, then select LOG ON. (No password
Needed).
2.) Select ABOUT tab.
3.) Verify that the SW versions is the same as what was recorded in
Torque wrench (available range: 12 Nm) with 5 mm hexagon bit
4.) Torque wrench with 5 mm hexagon bit to tighten the bolts to 12 Nm.