V1410 / V1412 User Manual
Mainstream/Sidestream CO
Messages
2
Message
Description
EtCO2 module is connected to the monitor
SENSOR WARM_UP
and is ready to operate.
Sample line is disconnected or disassembled from
SAMPLE LINE DISCONNECT
the module during measurement.
Measurement numeric value (or pressure) has
CHECK SAMPLING LINE
exceeded the allowable range, or the sample line
is not connected properly.
Respiration is not detected, or the sample line is
NO BREATHS DETECTED
disconnected from the patient.
CHECK ADAPTER
Verify sampling line is not occluded.
ZERO IN PROGRESS
Zero calibration is in progress.
Zero calibration is needed. Zero calibration is in
ZERO REQUIRED
progress.
SENSOR FAULTY
Sensor (EtCO2 module) is not connected.
Sample line is not connected to the sidestream
SENSOR DISCONNECT
module.
MESUREMENT MODE
EtCO2 measurement is in progress.
95