Vetland V1410 User Manual page 73

Veterinary patient monitor
Table of Contents

Advertisement

 VENOUS STAT Function
This function is used to easily find a patient's vein by controlling the flow of the vein by maintaining
constant pressure to the cuff.
1
2
3
To exit the VENOUS STAT mode, press NIBP.
 The VENOUS STAT function is only for patients whose veins are difficult to find. Avoid
using this function for normal patients.
Caution
 After using the VENOUS STAT function, set to AUTO or MANUAL for measurement.
 VENOUS STAT controls flow of the vein for 170 seconds for VET_ Large, and 85
seconds for VET_Small.
 VENOUS STAT is able to be set to between 50 – 200 mmHg (5 mmHg unit). The default
Note
setting is 100 mmHg.
Press the Menu button once and select, NIBP 
PRESSURE  VENOUS STAT and set the pressure
value.
Select NIBP  MODE  VENOUS STAT on the menu.
V1410 / V1412 User Manual