Electromagnetic Immunity - Vetland V1410 User Manual

Veterinary patient monitor
Table of Contents

Advertisement

V1410 / V1412 User Manual

Electromagnetic Immunity

Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short
Interruptions and
Voltage Variations
On Power Supply
Input Lines
IEC 61000-4-11
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
NOTE: U
is the A.C. main voltage prior to application of the test level.
T
108
IEC 60601
Compliance
Test Level
±6 kV Contact
±6 kV Contact
±8 kV Air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for
input/output lines
input/output lines
±1 kV differential
±1 kV differential
mode
±2 kV common
mode
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5cycle
for 0.5cycle
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycle
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycle
for 25 cycle
<5% U
T
(<95% dip in U
)
(<95% dip in U
T
for 5s
3 A/m
Electromagnetic
Level
Environment - Guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
±8 kV Air
covered with synthetic material,
the relative humidity should be at
least 30%.
Main power quality should be that
of a typical commercial or hospital
environment.
±1 kV for
Main power quality should be that
of a typical commercial or hospital
mode
environment.
<5% U
Main power quality should be that
T
of a typical commercial or hospital
)
T
environment. If the user of the
V1410 / V1412 requires continued
operation during power main
interruptions, it is recommended
40% U
T
that the V1410 / V1412 be
)
T
powered from an uninterruptible
for 5 cycle
power supply or a battery.
70% U
T
)
T
<5% U
T
)
T
for 5s
Power frequency magnetic fields
3 A/m
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.