V1410 / V1412 User Manual
 EtCO2 (End-Tidal CO2) – Optional
Microstream
- Measurement Mode: Microstream
- Range: 0 – 99mmHg
- Temperature: 0 – 50 ℃
- Accuracy:
Time
0 – 20 min
> 20 min
* (760 mmHg) level
- Respiration Measurement Range: 0 – 150 bpm
- Respiration Accuracy: 0 – 70 ±1 bpm
- Initial Time: Self test for 30 seconds after turning on the device.
- Sampling Value: 50 ±7.5 ml/min
- Period of Calibration: After 1 year or 4,000 hours of operation
- Response Time: 2.9 seconds after measurement (including rising time and delay time)
Mainstream
- Initial Time: less than 15 seconds (25
- CO
Measurement Range: 0 – 150 mmHg, 0 – 19.7%,
2
Resolution: 0 – 69 mmHg  0.1 mmHg
- CO
2
102
* CO
Density
2
0–38 mmHg
39–99 mmHg
0–38 mmHg
39–99 mmHg
71 – 120 ±2 bpm
121 – 150 ±3 bpm
Rising Time – 190 msec, Delay Time – 2.7 sec
0 – 20 kPa (at 760 mmHg)
70 – 150 mmHg  0.25 mmHg
Accuracy
±4 mmHg
±12%
±2 mmHg
±5% of reading + 0.08% for every
1 mmHg (above 38 mmHg)
C)