Accuracy: 0 – 40 mmHg  ±2 mmHg
- CO
2
41 – 70 mmHg  ±5% of reading
71 – 100 mmHg  ±8% of reading
101 – 150 mmHg  ±10% of reading
- Respiration Measurement Range: 0 – 150 bpm
- Respiration Accuracy: ±1 bpm
- Required Voltage: +0.5 VDC, 1.125 W (max 1.5 W)
- Temperature/Humidity: Operating Environment  0 – 45
- Waterproof: IPX-0
Sidestream
- Initial Time: less than 20 seconds (25
- CO
Measurement Range: 0 – 150 mmHg, 0 – 19.7%,
2
Resolution: 0 – 69 mmHg  0.1 mmHg
- CO
2
Accuracy: 0 – 40 mmHg  ±2 mmHg
- CO
2
- Respiration Measurement Range: 2 – 150 bpm
- Respiration Accuracy: ±1 bpm
- Required Voltage: +0.5 VDC ±5%
- Temperature/Humidity: Operating Environment  0 – 40
- Waterproof: IPX-0
Storage Environment  -40 – 70
C)
0 – 20 kPa (at 760 mmHg)
70 – 150 mmHg  0.25 mmHg
41 – 70 mmHg  ±5% of reading
71 – 100 mmHg  ±8% of reading
101 – 150 mmHg  ±10% of reading
More than 80 bpm ±12%
Storage Environment  -40 – 70
V1410 / V1412 User Manual
C, 10 – 90% RH
C, < 90% RH
C, 10 to 90% RH
C, < 90% RH
103