32. Respiration Measurement

Changes in the chest impedance resulting from the patient's respiration is measured and displayed as a
waveform and the values will appear on the screen.

How to Measure Respiration

1
2
Attach the electrodes for respiration measurement by referring to the figure below.
IEC
Color of
Lead
Electrode
R
Red
L
Yellow
F
Green
N
Black
C
White
<Sites for 3 Lead Attachment>
AAMI
Color of
Lead
Electrode
RA
White
Just below the clavicle (scapula) near the right shoulder
LA
Black
Just below the clavicle (scapula) near the left shoulder
LL
Red
Lower area of the left side of the abdomen
RL
Green
Lower area of the right side of the abdomen
Select the candidate lead for measurement among the chest
V
Brown
leads
Connect the ECG cable to the ECG connection
terminal of the measurement module.
Attaching Site
V1410 / V1412 User Manual