41. Product Specifications

 Measurement Parameters
3-channel ECG, NIBP, SpO2, 4 IBP, 4 Temp, EtCO2, or Respiration
 Processing Parameters
HR, NIBP (Systolic/Diastolic/Mean pressure), SpO2 Saturation, Pulse Rate, Respiration, IBP, Temp,
EtCO2
 Monitor Performances Specifications
- TFT Color LCD Display
- Waveform Display Method: Scroll
- Display Size: V1412 (12.1 inches) / V1410 (10.4 inches)
- Resolution : VGA (800 x 600)
- Sweep Speed: 6.25, 12.5, 25, 50 mm/sec
- Display waveform Time: 2.5 sec.
- Waveform Freeze: Function of waveform freeze
- Waveform Display: 3ch ECG, SpO2, 2 IBP, Respiration or EtCO2
- Waveform Traces: 9 Traces
- Measurement Value Display: Heart Rate, SpO2, Pulse Rate, NIBP, 4 IBP, 4 Temp, EtCO2.
 ECG Measurement
- ECG Wave: 3 leads (Standard - I, II, III) – 1 Channel
5 leads (Optional - I II, III, aVR, aVL, aVF, V) – 3 Channel, 7 Channel
- Input: Installed the circuit for protecting from the high frequency electric potential of ESU,
defibrillator and so on.
- Frequency Filter: High frequency noise remove filter / Install the 50/60Hz power noise remove filter.
- Input Impedance: Greater than 5 Mohm (50/60)
- Input for Initial Electric Current: Below 50 μA
- CMRR: Greater than 120dB
- Frequency Bandwidth: 0.5 – 40 Hz, 0.5 Hz – 80 Hz, 0.05 Hz – 40 Hz, 0.05 Hz – 80 Hz
- Gain: 2.5, 5, 10, 20, 30, 40 mm/mV, AUTO
- Calibration Waveform: 1 mVp-p old waveform, ±10%
- Input Range: Maximum Input Range: ±5mV