V1410 / V1412 User Manual

ECG Out Function (Optional)

The V1410 / V1412 provides ECG analog/pulse wave output from a rear stereo connector. It can be used
with defibrillator sync input or other medical devices.
How to Get ECG Output
1
2
Defibrillator Sync Signal
3
The ECG waveform out cable recommended by the manufacturer is described below.
58
ECG Output Port
ECG Analog Signal
Connect the cable to the rear radio connector of the
V1410 / V1412.
The blue waveform indicates an ECG analog signal. It can
be measured at the analog pin with a 1mV/mV unit.
The pink waveform indicates an ECG pulse detect signal.
It can be measured at a defibrillator sync pin with a 0–5V
unit.