Mainstream And Sidestream Co Calibration - Vetland V1410 User Manual

Veterinary patient monitor
Table of Contents

Advertisement

V1410 / V1412 User Manual
Mainstream and Sidestream CO
CO
Calibration should be executed at the following times:
2
– New patient
– When turning on/off the patient monitor
– When connecting or disconnecting EtCO2 module and sensor
– When a ZERO REQUIRED message appears during measurement.
CAUTION
EtCO2 calibration cannot be performed while in measurement mode. Before performing
calibration, remove the sample line from the patient and verify that a NO BREATHS
DETECTED message has appeared before proceeding.
1
2
92
Calibration
2
To set EtCO2 zeroing, press the Menu switch one
time on the Main screen and select
EtCO2  CALIBRATE.
To operate Calibration, select ZERO on the
CALIBRATE Menu.
(A ZERO IN PROGRESS message will be displayed
which means ZERO CALIBRATION is in progress).