V1410 / V1412 User Manual
Mainstream and Sidestream CO
CO
Calibration should be executed at the following times:
2
– New patient
– When turning on/off the patient monitor
– When connecting or disconnecting EtCO2 module and sensor
– When a ZERO REQUIRED message appears during measurement.
CAUTION
EtCO2 calibration cannot be performed while in measurement mode. Before performing
calibration, remove the sample line from the patient and verify that a NO BREATHS
DETECTED message has appeared before proceeding.
1
2
92
Calibration
2
To set EtCO2 zeroing, press the Menu switch one
time on the Main screen and select
EtCO2 CALIBRATE.
To operate Calibration, select ZERO on the
CALIBRATE Menu.
(A ZERO IN PROGRESS message will be displayed
which means ZERO CALIBRATION is in progress).