V1410 / V1412 User Manual
Microstream CO
The accuracy validation for the Microstream CO
measurement results are not accurate.
It is recommended to use 5% gas for calibration and the microstream filter line.
NOTE
Required Equipment
•
Calibration gas canister ( 5% CO2 medical)
•
Gas valve and tubing
•
Sample elbow or tee
•
Flow meter
88
Calibration
2
function is performed once a year or when the
2
(1) Connect the tube provided with the gas for
calibration to the gas barrel, and then connect it
to the filter line. Attach the gas/tube to the
V1410 / V1412 microstream port.
(2) Push the Menu button one time, select
EtCO2 Calibration CO2 and then select
CAS: 5.0%.
(3) Push the button on the 5% canister for
calibration to deliver the mixed gas.
(4) Push Menu button one time and select
EtCO2 Calibration Calibration ON.
(5) Check that the EtCO2 wave screen displays
"Calibration is in process. "
(6) When calibration is finished, remove the gas
tube.
(7) If a "CALIBRATION OK" message displays
on the screen after removing the gas tube, it is
properly done.