34. NIBP Measurement

 The NIBP blood pressure cuffs should only be used for blood pressure measurement. Proper
placement of the cuff and size selection is critical to achieving accurate readings. Use only a
Vetland NIBP cuff with the V1410 / V1412. Other NIBP cuffs may cause improper performance
due to tubing diameter or bladder compliance not meeting specifications.
WARNING
 Do not restrict or occlude the NIBP line.
 Store and use the hose properly.
 A patient's movement during measurement may cause inaccurate results.
 Do not use with any other respiration equipment. Choose the cuff type which is appropriate to
the patient.
 The NIBP module has an auto-calibration feature which is designed to sense the ambient
atmospheric air pressure every 0.5 seconds whenever the patient monitor is turned on.
CAUTION
(Attention: Any excessive or unnecessary pressure on the NIBP hose and NIBP cuff during
power up may influence the measurement value. In the case of an inaccurate auto-calibration
at power up, the unit will need to be powered off and back on to reset the auto calibration
correctly.
In the event a good blood pressure reading is not recorded on the initial cycle, the monitor will
attempt a second reading. If this fails, the unit will revert to standby mode. Check the BP line and
NOTE
cuff condition as well as size and placement.
V1410 / V1412 User Manual
67