Etco2 Measurement (Optional); Etco2 Measurement Screen - Vetland V1410 User Manual

Veterinary patient monitor
Table of Contents

Advertisement

39. EtCO2 Measurement (Optional)

Microstream, mainstream, and sidestream CO
microstream CO
module provides the function of EtCO2 and the function to measure respiratory CO
2
and respiration numbers using the small filter line based on a small lumen-type filter line. In the
microstream CO
type, the level is measured through the cannula (non-tubing type) attached to the nose
2
or a sampling line equipped in the respiratory circuit (tubing type).
The EtCO2 measurement utilizes a 50 ml/min sampling rate. The monitor should not be
attached to patients that may become distressed with a 50 ml/min reduced inspiratory
volume.
The sampling outlet of the V1410 / V1412 should be attached to an active scavenging
WARNING
system to evacuate the anesthetic gases discharged by the patient.
The gas inspiring vacuum pressure (negative pressure) of the gas elimination system should
not exceed the pump outlet standard of the V1410 / V1412, 1.0 mmHg. Excessive pressure
will display an OCCULSION message and it may damage the V1410 / V1412 pump. During
zeroing, the air elimination system should be operating.
CAUTION
Sampling lines and filters may become contaminated with patient bodily fluids. Use proper
protocols when handling such lines and accessories.
The monitor zeros the CO
sample lines to the CO
Range of Respiration Measuring: 0 – 150 bpm
For the best accuracy, a warm-up of about 20 minutes is required.
NOTE
It is optional to use 5% gas for calibration and the microstream filter line

EtCO2 Measurement Screen

No.
1
CO
Waveform
2
2
CO
Waveform Speed
2
3
CO
Waveform Size
2
4
Message Indicator
modules can be connected to the V1410 / V1412. The
2
sensor during the power-up self test. Do not allow connection of
2
source during the self test period.
2
Description
V1410 / V1412 User Manual
No.
Description
5
FiCO2
6
Respiration Measuring
7
EtCO2
2
87

Advertisement

Table of Contents
loading

This manual is also suitable for:

V1412

Table of Contents