Changing the NIBP Unit of Blood Pressure
1
Setting Mode
MODE
INITERVAL
STAT
PRESSURE
LEAKAGE
UNIT
Max Limit must be matched to the patient and set above anticipated systolic pressure to allow
for proper reading.
Caution
MAX LIMIT can be set between:
– VET_Large: 100 – 300 mmHg (5 mmHg unit)
– VET_Small: 70 – 300 mmHg (5 mmHg unit)
Note
Default setting of the MAX LIMIT is 280 mmHg.
Range
AUTO, MANUAL,
VENOUS STAT, and STAT
2, 3, 4, 5, 10, 15, 30 min,
1, 2, 4, 8, 12 Hours
5 –15 min (1 min interval)
MAX LIMIT, VENOUS STAT
mmHg, kPa
V1410 / V1412 User Manual
To change the unit of blood pressure, select UNIT
in the NIBP Menu and set the desired unit.
Description
Set the automatic measurement.
Set the interval of the automatic NIBP
measurement time.
Set the STAT MODE measurement time.
SCREEN: Set ON/OFF
MAX LIMIT: Limit value for pressure
VENOUS STAT: Pressure value of
VENOUS STAT MODE
Engineer Mode
Set the unit of blood pressure.
71