Installation Of Angle Rotors With A Hermetically Sealed Lid; Installation Of Accessories - Sigma 2-16KC Operating Manual

Refrigerated centrifuge
Hide thumbs Also See for 2-16KC:
Table of Contents

Advertisement

6.2.2.2

Installation of Angle Rotors with a Hermetically Sealed Lid

6.2.2.3

Installation of Accessories

Centrifugation with different tube sizes:
Working with different tube sizes is possible. In this case, however, it
is very important that opposite inserts are identical (see Fig. 6.2).
Centrifugation with low capacity: The tubes must be installed
symmetrically so that the buckets and their inserts are loaded evenly
(see Fig. 6.3).
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual
Screw the rotor cover onto the rotor and tighten it.
Lower the rotor with the cover onto the motor shaft .
Insert the rotor tie-down screw into the motor shaft. Tighten the
rotor tie-down screw with 5 Nm using the supplied rotor wrench
so that the spring washer assembly is compressed tightly.
The rotor can also be used without a cover.
Slightly grease the rotor and lid seals after cleaning.
The rotors can be installed or removed with a closed lid after
loosening the rotor tie-down screw.
Follow the safety instructions and hazard warnings in chapter 3!
1 Rotor tie down
screw
2 Lid
3 Lid seals
Fig. 6.1: Angel rotor with a hermetically sealed lid
Only use inserts that are suitable for the rotor (see 11.1 "Suitable
Accessories").
All four buckets of the swing-out rotor need to be installed when
spinning.
Always load the opposite inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
2-16KC
Using the Centrifuge
25