Using The Self-Test Utility; Requirements - Agilent Technologies Agilent E5250A User Manual

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Table 3-3
Executing Self-Test and Leak Test

Using the Self-Test Utility

Using the Self-Test Utility
The Self-Test Utility is a Self-Test program for the E5250A, and runs on the Agilent
4155/4156's built-in IBASIC controller.
Self-Test Utility can execute all Self-Test items listed in Table 3-1 and can also
execute the Leak Test, which checks leakage current of the E5250A with installed
cards. You execute the Self-Test items and Leak Test interactively (using softkeys)
from the 4155/4156.
This section explains the operation of the Self-Test Utility.

"Requirements"

"To Start the Self-Test Utility"
"To Execute Self-Test"
"To Execute Leak Test"
Requirements
The following equipment is required to execute the Self-Test Utility:
Required Equipment
Agilent
Model/Part No.
4155 or 4156
Semiconductor
Parameter Analyzer
10833A/B/C
GPIB Cable
E5250-17003
E5250A Program
Disk (furnished)
E5250A-301
Relay Test Adapter
(E5250-60002)
1250-0929
BNC Short Cap
3-10
Description
1
1
1
1
3 for each E5255A
(furnished with the
E5255A)
Agilent E5250A User's Guide, Edition 11
Qty.
Usage
Executing
Self-Test
Utility
Relay Test

Hide quick links: