Agilent Technologies Agilent E5250A User Manual page 201

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Syntax
Example
[:ROUTe]:CONNection:SEQuence
This command specifies the connection sequence mode, which is the open/close
sequence of the relays when connection route is changed from an existing
connection to a new connection. For details, refer to "Connection Sequence" in
Chapter 5.
The connection sequence is used only for cards that have been set to SROUte
connection rule. Refer to "ROUTe:CONNection:RULE".
At *RST, BBMake is selected.
[:ROUTe]:CONNection:SEQuence card_number,sequence
Parameter
card_number
sequence
OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,MBBR"
Agilent E5250A User's Guide, Edition 11
[:ROUTe]:CONNection:SEQuence
Explanation
card number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
NSEQ: No-Sequence mode
BBM: Break-Before-Make mode
MBBR: Make-Before-Break mode