NOTE
Installation

Self-Test

Self-Test
To confirm the specifications
The self-test and diagnostics checks the operation of the mainframe and the
modules. However they cannot confirm if the Agilent B2200 satisfies its
specifications.
For verifying the specifications, contact your nearest Agilent Technologies Service
Center. Trained service personnel will perform calibration (performance
verification).
It is recommended to perform calibration once a year at least.
The following procedure performs the self-test and diagnostics.
1. Turn on the Agilent B2200.
The controller test will be performed. If the Agilent B2200 fails the test, contact
your nearest Agilent Technologies Service Center.
2. Press the
key and the
Shift
the LCD.
SELF_TEST
RELAY_TEST
KEY
BEEPER
LED
PEN
GPIB
3. Move the cursor to the test item to perform, then press
4. To start the test, move the cursor to EXECUTE, then press
To display the previous test result, move the cursor to RESULT, then press
key.
For the details of each test, see "Selftest Menu" on page 3-34.
2-10
key in this order. Self-test menu will appear on
Menu
Controller test
Relay test
Front panel interface test
Beeper test
LED matrix test
Light pen test
GPIB test
Agilent B2200 User's Guide, Edition 2
key.
Enter
key.
Enter
Enter