Table 4-6
Setting up Measurement Environment
Measurement Cable Length
Guard Capacitances of E5250A Measurement Environment (Typical)
Connection
Agilent Model/Part No.
from SMU to
16494A-003 (for non-Kelvin)
E5250A
16494B-003 (for Kelvin)
16494C-001 (for 4142B)
E5250A and
E5252A (only 1 card is installed)
Plug-in Card
E5252A (more 1 card)
E5255A (only 1 card is installed)
E5255A(more 1 card)
from Card
16494A-001 (for non-Kelvin)
Output to
16494A-002 (for non-Kelvin)
Connector
Plate
16494B-001 (for Kelvin)
16494B-002 (for Kelvin)
16494C-001 (for 4142B)
16494C-002 (for 4142B)
16494D-001
16494D-002
16494E-003
from
Agilent Part No. 8121-1191
Connector
Plate to DUT
4-24
Cable
Length
80 cm
80 cm
1.5 m
1.5 m
3 m
1.5 m
3 m
1.5 m
3 m
1.5 m
3 m
3 m
X m
Agilent E5250A User's Guide, Edition 11
Guard
Capacitance
75 pF
90 pF
140 pF
145 pF
8 pF/extra card
180 pF
60 pF/extra card
125 pF
240 pF
140 pF
260 pF
140 pF
260 pF
125 pF
240 pF
240 pF
130 pF per m