Agilent Technologies Agilent E5250A User Manual page 285

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

To Execute HCI Test Program
1. Load the HCI test program (SP_MUX in this example), and press RUN function
key. Then the following message is displayed on the computer screen.
Enter directory name for saving data. (default: test_data)
2. Type the name of directory for saving HCI test data, and press
executing the program, test data is saved to the directory in the diskette. If you
press
key with no input, the test data will be saved in "test_data"
ENTER
directory.
After this operation, the Ib-Vgs measurement is automatically executed, and the
measurement results are displayed on the 4155/4156, as shown in Figure 9-6.
Also the program automatically extracts the gate stress voltage (Vgstr), which
will be used later for stress testing.
3. Remove the device used to determine Vdstr and Vgstr after the following
message is displayed.
"Connect HCI degradation test devices"
This device must be Device1, which was connected to the E5255A output port
numbers 1, 25, 49, and GNDU as shown in Figure 9-4.
4. Connect a new device in place of device removed by previous step. Then press
Continue function key to continue the program.
The program executes the leakage current tests, which determine the valid
devices to use for stress testing.
If the measured leakage currents are within the limit, following message is
displayed:
"Device No. = XX can be used"
If the measured leakage currents are not within the limit, following message is
displayed:
"Device No. = XX shall not be used"
Then, the initial characterization (before stress) is performed for all valid
devices.
Then, stress/interim characterization cycle is executed until stress termination
conditions are met as described previously. In each interim characterization,
Idlin, Gmmax, Vtext, and Vtci are determined. An example measurement
is shown in Figure 9-7.
Agilent E5250A User's Guide, Edition 11
Executing Sample Programs
HCI Measurement Program
key. After
ENTER
9-29