Agilent Technologies Agilent E5250A User Manual page 167

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

1. Measure R, L, C of your cable using the 4284A. Measurement setup is as
follows:
GUARD (C)
GROUND (E)
FORCE or SENSE (A)
Measurement
Parameter
R
L
C
a. See figure above.
b. Select 1 point in the frequency range.
c. For triaxial cable, connect B to F directly, and measure L
between A and E.
For coaxial cable, connect B to D directly, and measure L
between A and C. Ignore E and F.
2. Overwrite the R, L, C data in the program lines listed below. The data must be in
per meter values.
Program Line No.
1420
1430
1440
1450
Original program defines initial data by DATA statement in the "Rlc_data" block
as shown below:
1350 Rlc_data: !
1360
!
R [ohm]
:
:
1420
DATA 100.70E-3, 400.00E-9, 80.00E-12
1430
DATA 100.70E-3, 400.00E-9, 80.00E-12
1440
DATA 114.00E-3, 544.00E-9, 130.00E-12 !User Coax Cbl H
1450
DATA 114.00E-3, 544.00E-9, 130.00E-12 !User Coax Cbl L
:
:
Agilent E5250A User's Guide, Edition 11
Using the Capacitance Compensation Routine
Insulator
Triaxial Cable
Measurement
Measurement
Frequency
b
1 kHz to 1 MHz
PARALLEL
Defines R, L, C value of
Triaxial cable for C meter High terminal
Triaxial cable for C meter Low terminal
Coaxial cable for C meter High terminal
Coaxial cable for C meter Low terminal
L [H]
:
:
Programming the E5250A
GROUND (F)
GUARD (D)
FORCE or SENSE (B)
Measurement
Function
Terminals
A and B
SERIES
see note
A and C
C [F]
:
!User Triax Cbl H
!User Triax Cbl L
:
a
c
6-31