Electromagnetic Compatibility (Emc) And Electrostatic Discharge (Esd); What Is Emc; Compliance; Direction 5394227, 12 Logiq S8/Logiq E8 Service Manual - GE LOGIQ S8 Service Manual

Hide thumbs Also See for LOGIQ S8:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394227, 12
IRECTION
Section 1-9
Electromagnetic Compatibility (EMC) and
1-9-1

What is EMC?

Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
For applicable standards please refer to Chapter 2 in the Basic User Manual of the each system.
1-9-2

Compliance

The LOGIQ S8 conforms to all applicable conducted and radiated emission limits and to immunity from
electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient
requirements as mentioned in IEC 60601-1-2.
NOTE:
For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-9-3

Electrostatic Discharge (ESD) Prevention

DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
WARNING
WARNING
NECESSARY ESD PRECAUTIONS.
ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE POWER
CONNECTOR).
FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
DO NOT touch any boards with integrated circuits prior to taking the necessary ESD
WARNING
WARNING
precautions:
1.) When installing boards, ESD may cause damage to a board.
ALWAYS connect yourself, via an arm-wrist strap, to the advised ESD connection
point located on the rear of the system (to the right of the power connector).
2.) Follow general guidelines for handling of electrostatic sensitive equipment.
Risk of electrical shock: System must be turned off.
WARNING
WARNING
Avoid all contact with electrical contacts, conductors and components.
Always use non-conductive handles designed for the removal and replacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged
or isolated before making contact.
1 - 30
Section 1-9 - Electromagnetic Compatibility (EMC) and Electrostatic Discharge (ESD)
LOGIQ S8/LOGIQ E8 S
Electrostatic Discharge
M
ERVICE
ANUAL
(ESD)

Advertisement