Alignment and adjustment
Installing the tube adapter
•
Insert tube adapter to insertion depth 50
50mm
mm
•
Corrections from min. 40 mm to max. 55
mm (with integrated hard stop) insertion
depth are possible
•
Pre-fasten both screws lightly at the clamp
(only valid for the basic alignment, not
for the static or dynamic alignment!)
Certification course | C-Leg 4 | Global Academy |
For safety reasons,
always insert the tube at
least to 40 mm!
25.09.2019
| © Ottobock
61