Gait cycle
Stance release
Knee
extended
•
The C-Leg 4 monitors the gait cycle in real time
•
Stance release is independent of body weight
changes and ground condition
•
Main criteria for stance release:
-
Extended (hyperextended) knee joint
AND
-
Forward inclination of the shank
AND
-
Forward rotation of the shank
Due to the IMU walking
backwards safely is possible!
Certification course | C-Leg 4 | Global Academy |
25.09.2019
| © Ottobock
29