User Instructions
Safety modes – Drained battery
•
This error is displayed with 10 beeps and vibration
signals
•
The safety mode flexion resistance can be set
individually by the CPO in C-Soft Plus and starts
gradually in the case of a drained battery (within
max. 10 seconds)
•
With a low resistance, the joint has to be actively
secured upon heel impact
In the case of a drained battery, the
prosthesis switches back to basic mode
after the battery is charged!
Certification course | C-Leg 4 | Global Academy |
25.09.2019
| © Ottobock
135