Alignment and adjustment
Using C-Soft Plus for bilateral fittings
If just one BionicLink PC is available or a second BionicLink PC is not be
detected by the computer:
L
•
•
•
•
•
R
•
Certification course | C-Leg 4 | Global Academy |
Open the DataStation twice and
select C-Soft Plus in each window
Connect to the first knee joint with
the first instance of C-Soft Plus
and adjust the settings
Do not close the whole job, but
instead disconnect the knee joint
from the software (Tab
"Connection" und Button
"Disconnect")
Connect the second knee joint to
the second instance of C-Soft Plus
Make the further adjustments
using the same procedure
Save both jobs when completed
25.09.2019
| © Ottobock
106