User Instructions
Storage and bleeding air
•
During extended non-vertical storage of the joint,
air may accumulate in the hydraulics
•
Ideal storage position: vertical with extended knee
head
•
This is noticeable through sounds and irregular
damping behavior
Automatic air bleeding
mechanism after 10 – 20 steps!
Certification course | C-Leg 4 | Global Academy |
25.09.2019
| © Ottobock