NOTE

To Change Connection Rule

1. Press the
key. The LCD displays the present setting (FREE or SINGLE).
Rule
2. Press the arrow key to select the connection rule, then press the
press the
key to cancel this operation.
Exit

To Change Connection Sequence

1. Press the
key and the
Shift
(Break_Before_Make, Make_Before_Break, or No_Sequence).
2. Press the arrow key to select the connection sequence, then press the
Or press the
key to cancel this operation.
Exit

To Control Switch Condition

If you use the light pen, you do not need the following instruction to control the
switch condition.
1. If you are NOT watching the orange-blink-LED on the LED matrix, press the
key. The blink-LED will appear.
Open/Close
2. Press the arrow keys to move the blink-point.
3. Press the
Open/Close
The green-LEDs indicate the switches that are closed now.
4. Repeat 2 and 3 until the switch setup is completed.
Blinking of LED will be stopped by pressing the front panel key other than the
,
, or arrow keys.
Open/Close
Fast
When Kelvin cables are connected to Outputs
For the Kelvin outputs, the input-output paths must be controlled as shown below.
For the input couple port:
Control the switches to connect the input couple port to the Kelvin output.
For the single input port:
Control the switches to connect the input port to two ports of the Kelvin output.
If you do not comply with this note, leakage current between the Kelvin paths will
cause measurement error.
Agilent B2200 User's Guide, Edition 2
key. The LCD displays the present setting
Rule
key to open/close the switch specified by the blink-LED.
Front Panel Operation
Operation
key. Or
Enter
key.
Enter
3-5