B2200 Channel Related Errors - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

B2200 Channel Related Errors

Error
Number
2000
Invalid card number
Wrong card number is specified in card number or channel list
parameter. Make sure card is properly installed in the correct
B2200 slot.
2001
Invalid channel number
Wrong channel number is specified in the channel list. Confirm the
channel number, the card configuration, and the configuration
mode of the B2200.
2002
Unsupported Module
2003
Unsupported Configuration
2006
Command not supported on this card
Command was used that is not supported by the card.
2009
Too many channels in channel list
Too many channels are specified in the channel list (maximum
120).
2011
Empty channel list
No channel list is specified.
2012
Invalid channel range
Wrong channel list is defined. Confirm the syntax of the channel
list, the card configuration, and the configuration mode of the
B2200.
Agilent B2200 User's Guide, Edition 2
B2200 Specific Error Messages
Error Message and Description
Error Messages
7-11