Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Mating Connectors

Figure 2-3 provides interconnection data for the General Purpose
Interface Bus.

Figure 2-3 General Purpose Interface Bus Connection

Logic levels
The interface bus logic levels are TTL compatible, that is, the true (1)
state is 0.0 to +0.4 Vdc and the false (0) state is +2.5 to +5.0 Vdc.
Programming
Refer to Chapter 6, "Remote Operation".
and Output data
format
Mating connector Agilent 1251-0293
Mating cables
Agilent 10833A, 1 meter (3.3 ft.), Agilent 10833B, 2 meters (6.6 ft.)
1
Agilent 10833C, 4 meters (13.2 ft.), Agilent 10833D, 0.5 meter (1.6 ft.)
available
The maximum accumulative length of connecting cable for an interface
CAUTION
bus system is 20 meters (65.5 ft.)
A general purpose interface bus system may contain no more than 2
meter (6.6 ft.) of connecting cable per instrument.
1. The Agilent 10834A adapter is required with these cables to provide additional clearance for the
cable connections.
2-6 11713A Operating and Service Manual
Mating connectors used with the driver are the 24-pin GPIB male,
the 12-pin Viking Industries connector, and the banana type male.